6 - 31

IM 11G02N01-01E

5) On the parameter hold screen that

appears, move the cursor next to the Ch
(component) you want to make the
setting by pressing the

or the

key,

and then press the

ENT

key.

6) The value is highlighted, indicating that

the value can be changed. Change the
value by pressing the

or the

key,

and then move the cursor to the right by
pressing the

key.

7) After the value is changed, press the

ENT

key.

8) Press the

ESC

key to return to the

parameter setting screen.

The setting is expressed in % against the range
for both ranges.
When 0 to 1000 ppm is selected as the range,
for example, if 10% FS is selected as hold
setting, the output equivalent to 100 ppm is
output and held irrespective of the measure-
ment value at that time.

Meaning of setting

Instantaneous value display of the measurement cannot be held. (Output only can be held.)

If set value is selected for hold, instantaneous O

2

correction value is calculated and held

based on the set value.

Range identification contact output cannot be switched even if the range is switched during
the hold.

Description of setting

ESC

End of Hold Setting

Parameter Setting screen

ENT

(

)

ENT

(

)

ENT