Yokogawa IR400 User Manual

Page 26

Advertising
background image

3 - 13

IM 11G02N01-01E

Terminal block 3 <TN3>

Between 1–4:

For internal connection. Must
not be wired. (Must not be used
as junction terminal.)

Between 5–6:

Remote hold input. No hold
when open. Output hold when
short-circuited.

For details, refer to “Item 6.7
Parameter setting, Output Hold”.

Between 7–8:

Average value reset input. short-
circuitting the contact input (for
at 1.5 sec min.) resets O

2

average

and O

2

corrected average simulta-

neously. Opening it restarts the
average value.

For details, refer to “Item 6.7
Parameter setting, Average Value
Resetting”

Between 9–10:

Automatic calibration remote
start input

After shorting for 1.5 sec. or
more, automatic calibration is
started by the opening input
whether the automatic calibration
setting is ON/OFF.

For details, refer to “Item 6.4
Setting of auto calibration”

Between 11–12:

Ch5 range identification contact
output

Between 13–14:

Ch4 range identification contact
output

Between 15–16

Ch3 range identification contact
output

Between 17–18:

Ch2 range identification contact
output

Between 19–20:

Ch1 range identification contact
output

Action of range identification signal

Range identification contact is
conductive at low range and open
at high range.

Terminal block 3

<TN3>

11

12

13

14

15

16

17

18

19

20

1

2

3

4

5

6

7

8

9

10

(M3.5 screw)

Ch4 range identification
contact output (RNG_IDCh4)

Ch5 range identification
contact output (RNG_IDCh5)

Ch3 range identification
contact output (RNG_IDCh3)

Ch2 range identification
contact output (RNG_IDCh2)

Ch1 range identification
contact output (RNG_IDCh1)

Auto calibration

remote start

input (R_CAL)

Average value reset

input (RESET)

Remote hold input

(R_HOLD)

Unassigned

Unassigned