Specifications, 1 genenal specifications, 1 general specifications – Yokogawa IR400 User Manual

Page 96: Standard specifications

Advertising
background image

9 - 1

IM 11G02N01-01E

1. Standard specifications

Measurable gas components and measuring range:

• Max. 5 components measurement

including O

2

.

• Measuring range ratio

1:5 (O

2

)

1:25

(except for O

2

)

• For measurable components and

possible combinations of measuring
ranges, refer to Tables 1-(1) to (7).

Principle of measurement:
NO, SO

2

, CO

2

, CO, CH

4

;

Non-dispersion infrared-ray absorption
method
Single light source and double beams
(double-beam system)

O

2

; Paramagnetic O

2

sensor (built in) or zir-

conia O

2

sensor (externally installed)

Measured value indication:

Digital indication in 4 digits
(LCD with back light)
• Instantaneous value of each component
• Instantaneous value after O

2

correction

(only in NO, SO

2

, CO measurement

with O

2

)

• Average value after O

2

correction

(only in NO, SO

2

, CO measurement

with O

2

)

• O

2

average value

Analog output signals:

* Inputs/outputs of analog signals are

possible by combining with the input/
output terminal module.

4 to 20mA DC or 0 to 1V DC,
non-isolated output ; 12 points max.
max.load

550

. for 4 to 20 mA DC

min.load

100k

. for 0 to 1V DC

* Refer to Table 2, for the channel No. of

displayed values and analog output
signals.

Analog input signal:

For signal input from externally installed
O

2

sensor.

Signal requirement;
(1) Signal from Zirconia O

2

sensor (when

ZX8D

ء

C or ZX8D

ء

D)

(2) 0 to 1V DC from an O

2

sensor Input

section is not isolated. This feature is
effective when an O

2

sensor is not

built in.

(Depend on O

2

input signal, measured

concentration indication and O

2

conver-

sion.)

9.

SPECIFICATIONS

9.1

General specifications

Minimum range

Maximam range

CO

2

CO

CH

4

O

2

(built in)

O

2

(External
Zirconia)

0 – 50ppm

0 – 100vol%

0 – 100vol%

0 – 100vol%

0 – 25vol%

0 – 200ppm

0 – 5vol%

0 – 20ppm

0 – 25vol%

0 – 5vol%

NO

SO

2

0 – 50ppm

0 – 5000ppm

0 – 10vol%

0 – 50ppm

Relay Contact output:

1a contact (250V AC/2A, resistive load)

Instrument error, calibration error,
range identification, auto calibration
status, pump ON/OFF, peak alarm.

1c contact (250V AC/2A, resistive load
selectable 6 outputs)

High/Low limit alarm contact output.
Power disconnection alarm.

* All relay contacts are isolated mutually

and from the internal circuit.

Contact input: No-voltage contact (ON/0V, OFF/5V

DC, 5mA flowing at ON)

Remote range switch, auto calibration
remote start, remote holding, average
value resetting, pump ON/OFF

Isolated from the internal circuit with
photocoupler. Contact inputs are not iso-
lated from one another.

Transmission output:

Solenoid valve drive signal for automatic
calibration.
Transistor output (100mA or less)

Power supply: Voltage rating

; 100V to 240V AC

Allowable range ; 85V to 264V AC
Frequency

; 50Hz/60Hz

Power consumption; 250VA max.
Inlet

; Conform to EN60320

Protection Class I type

Operation conditions:

Ambient temperature ; -5˚C to 45˚C
Ambient humidity

; 90% RH max.,

non-condensing

Storage conditions:

Ambient temperature ; -20˚C to 60˚C
Ambient humidity

; 100% RH max.,

non-condensing

Dimensions (H

ЧЧЧЧЧ

W

ЧЧЧЧЧ

D):

Analyzer main unit;
177 x 483 x 578mm (Excluding handles.)
Input/output terminal module;
164 x 318 x 55mm

Mass:

Approx. 22 kg (only Analyzer)

Finish color:

Front panel ; Off-white (Munsell 10Y7.5/

0.5 or equivalent)

Casing; Steel-blue (gray)

Enclosure:

Steel casing, for indoor use

Material of gas-contacting parts:

Gas inlet/outlet; SUS304
Sample cell; SUS304,chloroprene rubber
Infrared-ray transmitting window; CaF

2

O

2

sensor sample cell : SUS316

Internal piping; Toaron, Teflon

Gas inlet/outlet: Rc1/4 or 1/4NPT internal thread
Purge gas flow rate: 1L/min ( when required)

Installation altitude: 2000 m or less
Safety and EMC conforming standards:

Safety:

EN61010-1
Pollution degree 2 (Note)

Installation category II (Note)

Note: Installation category, called over-voltage

category, specifies impulse withstanding
voltage. Category II is for electrical equip-

ment.

Pollution degree indicates the degree of

existence of solid, liquid, gas or other

inclusions which may reduce dielectric
strength. Degree 2 is the normal indoor

environment.

EMC:

EN61326
EN61000-3-2
EN61000-3-3