<Gas flow time> setting

(1) Press the

ENT

key in a state where the

cursor is placed next to “Flow Time,” and
the flow time setting screen shown at right
appears.

(2) On the flow time setting screen that

appears, move the cursor to the gas you
want to change the setting by pressing the

or the

key, and then press the

ENT

key.

(3) The highlighted value can be changed.

Change the value by pressing the

or

the

key, and then move the cursor to

the right by pressing the

key.

(4) After changing the value, press the

ENT

key.

(5) Press the

ESC

key to return to the

automatic calibration setting screen.

Note) Only the Chs used are displayed on

this screen. The Ex. time is the
output signal hold extension time
after the completion of calibration.
It is valid only when the hold setting
is set to “ON.” The Ex. time set
here is also the hold extension time
at the time of manual calibration.

MODE

(

)

ENT

ENT

End of Gas Flow Time Setting