2 forced run/stop of calibration, 1 execution of auto calibration*only once – Yokogawa IR400 User Manual

Page 56

Advertising
background image

6 - 17

IM 11G02N01-01E

(

)

ENT

(

)

ENT

6.4.2 Forced run/stop of auto calibration

Auto calibration can be performed just once or forcibly stopped while the calibration is
performed.

6.4.2.1 Execution of auto calibration (only once)

(1) Display the User mode screen. Move the

cursor to “Setting of Auto Calibration”
by pressing the

or the

key, and

then press the

ENT

key.

(2) In the “Setting of Auto Calibration” item

selection screen that appears, point the
cursor to “Auto Calibration Run” by
pressing the

or

key. Press the

ENT

key.

(3) “Run” is highlighted, displaying a mes-

sage to confirm the execution of auto
calibration. Press the

ENT

key to execute

the auto calibration forcibly, and press
the

ESC

key to cancel.