6 - 16

IM 11G02N01-01E

Remote start

Whether the auto calibration is set at ON or OFF, an auto calibration is available by keeping the
remote start input closed for at least 1.5 seconds.

Remote start input

Open

Closed (keep at least 1.5 sec.)

Start Time
Cycle
Flow Time

ON/OFF

SUN
1
Zero
Ch1 Span
Ch2 Span
Ch3 Span
Ch4 Span
Ch5 Span
EX. time
ON

12:00
day
350 sec
350 sec
350 sec
350 sec
300 sec
300 sec
300 sec

(An example of “Ch1: through Ch5: enable”, as given in Item 6.2.4 “Auto Calibration
Components/range”)

Flow time

: Auto calibration

Sunday
12:00

Cycle

Zero
calibration

350sec

350sec

Ch1
Span
calibration

350sec

Ch2
Span
calibration

Ch3
Span
calibration

Ch4
Span
calibration

Monday
12:00

Tuesday
12:00

Example

In case where auto calibration is carried out at the above setting.

Ch5
Span
calibration

350sec

300sec

300sec

Replace-
ment time

300sec

Cycle

: 1 to 99 hours or 1 to 40 days (initial value 7days)

Flow time

: 60 to 900 sec

(initial value 300sec)

Setting range

When an auto calibration starts, the measurement screen automatically appears.

Any key operation other than key lock ON/OFF and “Stop Auto Calibration” (see

Item 6.4.2) is not permitted during auto calibration. “Stop Auto Calibration” cannot

be performed with the key lock to ON. To cancel auto calibration forcedly, set

the key lock to OFF and then execute “Stop Auto Calibration”.

Turn on the power again after it is turned off (including the case of power

failure) at the time set as the next start time in auto calibration, and then

repeat it in the set cycle.

Caution


Auto calibration status contact output is closed during auto calibration (including Ex. time), and is open in
other cases.