6 - 4

IM 11G02N01-01E

ENT

End of Calibration

Concentration Setting

NOx, SO

2

, CO

2,

CO, CH

4

, external O

2

measurement and buit-in paramagnetic O

2

sensor

Span gas: 1 to 105% of full scale
(Full scale (FS) is the same as each range
value.)

External Zirconia O

2

measurement

The setting cannot be performed beyond the range.

Zero gas: 5 to 25 vol%
Span gas: 0.01 to 5 vol%

Setting range of values

Cursor for setting value

(5) In the “Calibration Concentration Selection”

screen that appears, select any concentration
item you want to set by pressing the

,

key.

Then press the

ENT

key, and the selected

value is highlighted.

(6) In the “Calibration Concentration Value

Setting” screen that appears, enter calibration
gas concentration values (zero and span).
For value entry, press the

or

key, and

a 1-digit value increases or decreases. By

pressing the

, the digit moves.

After setting, save the entry by pressing the

ENT

key. The saved value becomes valid

from the next calibration process.

Note) Enter settings that correspond to each

range. If zirconia type is used as O

2

sensor, select 21.00 for the field of Zero
(when air is used), and select the
concentration listed on the cylinder if the
air contained in a cylinder is used.

To close the calibration concentration value
setting process or cancel this mode midway,
press the key.
A previous screen will return.

To close the setting

ESC

ENT