6.4 Setting of auto calibration

6.4.1 Auto calibration

Auto calibration is automatically carried out at the time when zero calibration and span calibration
are set.

Before changing the setting of auto calibration, set the ON/OFF to OFF.

1.

During measurement, press the

MODE

key

to display the User mode.

2.

Point the cursor to “Setting of Auto
Calibration” by pressing the

or

key. Press the

ENT

key.

3.

In the “Setting of Auto Calibration”
screen that appears, point the cursor to
any item you want to set by pressing the

or

key. Press the

ENT

key.

4.

In the “Auto Calibration Parameter
Setting” screen that appears, perform the
value entry or the setting. For the value
entry or setting change, use the

or

key. To change the setting, use the

key to move the cursor to the right.

After setting, press the

ENT

key, and

setting of auto calibration is carried out .

(

)

ENT

MODE

(

)

ENT

ENT

End of Aoto Calibration Setting

To close the "Setting of Auto calibration" or cancel

this mode midway, press the

ESC

key.

A previous screen will return.

To close "Setting of Auto calibration"

ESC

Description of setting items

• Start Time : Setting at the first calibration

(day of the week, hour, minute)

• Cycle

: A period between the start time of one

calibration and another
(unit : hour/day)

• Flow Time : The time required for replacement by

calibration gas
Time required for replacement of sample
gas after the calibration is completed
(Set by calibration gas. See the next
page.)

• ON/OFF

: ON/OFF of auto calibration