3 setting of calibration range – Yokogawa IR400 User Manual

Page 46

Advertising
background image

6 - 7

IM 11G02N01-01E

(

)

ENT

(

)

ENT

MODE

(

)

ENT

(1) During measurement, press the

MODE

key to

display the User mode.

(2) Point the cursor to “Calibration

Parameters” by pressing the

or

key. Press the

ENT

key.

(3) In the “Calibration Parameters” screen

that appears, point the cursor to “About

Calibration Range” by pressing the

or

key. Press the

ENT

key.

(4) In the “Calibration Range Ch Selection”

screen that appears, point the cursor to the

Ch you want to set by pressing the

or

key. Press the

ENT

key.

6.2.3 Setting of calibration range

This mode is used to set if the range of each CH (component) at the calibration (manual calibration
or auto calibration) should be calibrated with a single range or 2 ranges.