3 alarm setting, 1 setting of alarm values – Yokogawa IR400 User Manual

Page 50

Advertising
background image

6 - 11

IM 11G02N01-01E

(

)

ENT

MODE

(

)

ENT

Note

Set the values so that H-limit value > L-limit
value and that (H-limit value

-

L-limit value)

> hysteresis.

6.3 Alarm setting

6.3.1 Setting of alarm values

The High/Low limit alarm output setting for the measured concentration and power off alarm
(alarm 6 only) setting can be made during measurement. Arbitrary 6 alarm contact outputs can be
used.

To change alarm setting, set the alarm ON/OFF setting to OFF, and then change the value.

(1) During measurement, press the

MODE

key

to display the User mode.

(2) Point the cursor to “Alarm Setting” by

pressing the

or

key. Press the

ENT

key.

(3) After the alarm No. selection screen has

appeared, point the cursor to the Alarm
No. you want to set by pressing

or

key .

Press the

ENT

key.

(4) After the alarm item selection screen has

appeared, operate the

or

key until

the cursor is aligned with a desired item
and press the

ENT

key.

(

)

ENT

Advertising