3.2 Installation

3.2.1 Installation of analyzer main frame

Installation methods for the analyzer main unit are divided into 2 types;

Note 1 Check and maintenance of the analyzer main unit may be carried out with the top cover detached.

The guide rail method may be used if a space accessible for maintenance is provided at the top of
the main unit. If maintenance space is not provided specially, it is recommended to use the slide
rail method.

Recommended slide rail: Product No.: 305A-24 manufactured by Accuride International Co.

Note 2 For 19 inch rack mounting, the weight of the analyzer is supported with the bottom of the case

(with the side of the case in case of slide rail method). For mounting dimensions of the slide rail,
see “Item 9.3 External diagram”.

Don’t install the analyzer at a place which is exposed to direct sunlight.

The analyzer should be installed at a place where ambient temperature is within -5 to 45˚C, and

temperature fluctuation during use is minimum.

450 or more

465

101.6

M6

Slide rail
(Supports mass)

Slide rail
(Supports mass)

Type

External dimensions

Mounting dimensions

Mounting method

19 inch rack

mounting guide

rail method

19 inch rack

mounting slide

rail method

429

177

483

(Unit : mm)