Yokogawa IR400 User Manual

Page 49

Advertising
background image

6 - 10

IM 11G02N01-01E

(5) The cursor next to the range of the

selected Ch (component) is highlighted.
Select the range to be calibrated mainly
by pressing the

or the

key.

(6) Then press the

ENT

key, and calibration is

performed in the selected range.

Auto calibration and the manual calibration

of the component with which “AR” has

been selected as range switch mode are

performed in the range selected here. In this

case, once the calibration is started, the range

is automatically switched, and on completion

of the calibration, the original range is

resumed.

The range identification contact is inter-

locked with the range after the switch.

However, if the hold setting is set to “ON,”

the contact status before calibration is main-

tained.

To close "Auto Calibration
Component/range" setting

(7) Press the

key in the state described in

(5), and the highlight is switched between
“enable” and “disable” auto calibration.

(8) Select “enable” of “disable” by pressing

the

or the

key.

(9) Then press the

ENT

key.

Press the key to exit automatic calibra-

tion component/range setting, and the pre-

vious screen appears.

To close the setting

ESC

Operation by setting

Auto calibration is performed under the following rules.

1. Zero calibration is performed at the same time, for the Ch (component) with which “enable”

is selected at the time of auto calibration and auto zero calibration.

2. Span calibration is performed in the order from smallest Ch No., for the Ch (component)

with which “enable” is selected at the time of auto calibration.

ZERO calibration on a

uto calibration and auto zero calibration of the component with

which “enable” is selected are performed in batch irrespective of the description in

“6.2.2 Setting of manual zero calibration.”

Note

(

)

ENT

End of Auto Calibtation

Range Setting

(

)

ENT

End of Auto Calibtation

Component Setting