2 span calibration – Yokogawa IR400 User Manual

Page 80

Advertising
background image

6 - 41

IM 11G02N01-01E

When “both” from “Calibration Range” of the
Calibration Setting mode is set, span calibra-
tion is performed together with 2 Ranges.

Caution

To close the “Span Calibration” or cancel
this mode midway, press the key.
A previous screen will return.

To close "Span Calibration"

ESC

(1) Press the

SPAN

key on the Measurement

screen to display the Manual Span Calibra-

tion screen.

(2) Select CH (component) to be calibrated by

pressing the

or

key and press the

ENT

key. The calibration gas is supplied.

To Measurement screen after

executing Manual Span Calibration

ENT

SPAN

(3) Wait until the indication is stabilized in the

state where the calibration gas is supplied.
After the indication has been stabilized,
press the

ENT

key. Span calibration of

Range selected by the cursor is performed.

Note: For the Ch (component) for

which“AR” is selected in “6.1.1
Setting range switch mode,” the
cursor automatically moves to the
range selected in “Setting of auto
calibration component/range”
(6.2.4), and calibration is carried
out within that range.

(

)

ENT

6.9.2 Span calibration

It is used to perform a span point adjustment. Supply calibration gas with concentration set to the
span value to perform the span calibration. For the span calibration gas for the NO

X

, SO

2

, CO

2

,

CO measurement, use the standard gas with a concentration of 90% or more of the range value.
For the span calibration gas for the O

2

measurement, use the standard gas with a concentration of

90% or more of the range value when measuring with the built-in O

2

sensor, and use the standard

gas of 1to 2 vol% when measuring with an external zirconia O

2

sensor.