3 - 12

IM 11G02N01-01E

(6) Description on terminal block

Terminal block 1

<TN1>

11

12

13

14

15

16

17

18

19

20

1

2

3

4

5

6

7

8

9

10

(M3.5 screw)

Ch9 output
(Ch9_OUT)

Ch10 output
(Ch10_OUT)

Ch8 output
(Ch8_OUT)

Ch7 output
(Ch7_OUT)

Ch6 output
(Ch6_OUT)

Ch1 output

(Ch1_OUT)

Ch2 output

(Ch2_OUT)

Ch3 output

(Ch3_OUT)

Ch4 output

(Ch4_OUT)

Ch5 output

(Ch5_OUT) +

+

+

+

+

+

+

+

+

+

+

Terminal block 2

<TN2>

11

12

13

14

15

16

17

18

19

20

1

2

3

4

5

6

7

8

9

10

(M3.5 screw)

Ch4 remote range
changeover input
(R_RNG_Ch4)

Ch5 remote range
changeover input
(R_RNG_Ch5)

Ch3 remote range
changeover input
(R_RNG_Ch3)

Ch2 remote range
changeover input
(R_RNG_Ch2)

Ch1 remote range
changeover input
(R_RNG_Ch1)

Ch12 output

(Ch12_OUT)

Ch11 output

(Ch11_OUT)

Unassigned

Unassigned

O

2

sensor input

(O

2

_IN)

Note 1

+

+

Note 1:

For external O

2

sensor input.

Terminal block 1 <TN1>

Terminal block for analog output (non-isolated output)

Between 1–2:

Ch5 output

Between 3–4:

Ch4 output

Between 5–6:

Ch3output

Between 7–8:

Ch2 output

Between 9–10:

Ch1 output

Between 11–12:

Ch10 output

Between 13–14:

Ch9 output

Between 15–16:

Ch8 output

Between 17–18:

Ch7 output

Between 19–20:

Ch6 output

Terminal block 2 <TN2>

Between 1–2:

O

2

sensor input

(For input of Fuji’s zirconia
oxygen sensor or externally
oxygen sensor. Must not be used
unless external O

2

sensor is

provided.)

Between 3–4:

Ch12 output

Between 5–6:

Ch11 output

Between 7–10

For internal connection. Must not
be wired. (Must not be used as
junction terminal).

Between 11–12:

Ch5 remote range switch input

Between 13–14:

Ch4 remote range switch input

Between 15–16:

Ch3 remote range switch input

Between 17–18:

Ch2 remote range switch input

Between 19–20:

Ch1 remote range switch input

Action of remote range switch

High range is selected when open.
Low range is selected when short-
circuited. For details of action, see
“Item 6.1 Switch of range.”

Advertising