9 calibration, 1 zero calibration – Yokogawa IR400 User Manual

Page 79

Advertising
background image

6 - 40

IM 11G02N01-01E

ENT

To Measurement screen after

executing Manual Zero Calibration

ZERO

6.9.1 Zero calibration

It is used for zero point adjustment. For zero
calibration, gas suited for an application
should be used according to “(3) Standard gas
in Item 3.3 Sampling.”

(1) Press the

ZERO

key on the Measurement

screen to display the Manual Zero Cali-

bration screen.

(2) Select the Ch (component) to be

calibrated by pressing the

or

key.

After selection, press the

ENT

key, and

zero gas will be supplied.

(3) Wait until the indication is stabilized

with the zero gas supplied. After the

indication has been stabilized, press the

ENT

key. Zero calibration in range

selected by the cursor is carried out.

Note: For the Ch (component) for

which “AR” is selected in “6.1.1
Setting range switch mode,”
the cursor automatically moves
to the range selected in “Setting
of auto calibration component/
range” (6.2.4), and calibration is
carried out within that range.

For the Ch (components) that is set to
“both” in the “Zero Calibration” of the
Calibration Setting mode, zero calibration
is also carried out at the same time.

Caution

To close the “Zero Calibration” or

cancel this mode midway, press the

ESC

key. A previous screen will return.

To close "Zero Calibration"

ESC

(

)

ENT

6.9 Calibration