IM 11G02N01-01E

In case of Error No. 5 and No. 7

Error log file

If error occurs, the history is saved in an error log file. The error log file exists in the mainte-
nance mode.

Error log screen

* Up to 14 errors can be saved in the error history; the oldest error will be deleted one by one every

time a new occurs.

* If the power supply is turned OFF, the contents in the error log file will not be lost or damaged.

Deletion of error history

Press the

ENT

key on the above screen, and the “Error Log Clear” will be inverted. Further

pressing the

ENT

key will clear the error history.

• Pressing delete the error display.

Calibration is continued. Unless another calibration
error occurs, calibration is carried out to the end, the
Measurement screen returns.

ENT

ENT