Section 7. Installation

158 

 

 

Calibration Report for Pressure Transducer


Parameter

Measurement
Before Zero

Measurement
After Zero

Piezometer Output (digits)

8746

0

Piezometer Temperature (°C)

21.4

0

Barometer Pressure (mb)

991

0

 

1. Send CRBasic example FieldCal() Zero Basis Demo Program

(p. 158)

to the

CR1000.

2. To simulate the pressure transducer in zero conditions:

Digits_Measured is set to 8746 automatically

Temp_Measured is set to 21.4 automatically

BP_Measured to 991

3. To simulate the calibration, change the value in variable CalMode to 1 to start

calibration. When CalMode increments to 6, zero calibrations are complete.
Calibrated offsets will equal 8746 digits, 21.5 C°, and 991 mb.

CRBasic Example 28. FieldCal() Zero Basis Demo Program 

'FieldCal zero basis demonstration program

Public

Pressure

Public

VW(1,6)

Public

Equation_Parameters(3)

Alias

Equation_Parameters(1) = Digits_Measured

Alias

Equation_Parameters(2) = Temp_Measured

Alias

Equation_Parameters(3) = BP_Measured

Public

Offset(3)

Alias

Offset(1) = Digits_Offset

Alias

Offset(2) = Temp_Offset

Alias

Offset(3) = BP_Offset

Public

LoadResult, CalMode

Public

AVWRC

Const

GageFactor = 0.01664

Const

Temp_K = -0.00517

BeginProg

'Load the calibration constants stored in the CAL file after a zero is performed

LoadResult = LoadFieldCal(False)

Scan

(1,Sec,1,0)