4 pt100 in four-wire full-bridge, Figure 80: pt100 in three-wire half-bridge – Campbell Scientific CR1000 Measurement and Control System User Manual

Page 261

Advertising
background image

Section 7. Installation

261

 

Figure 80: PT100 in three-wire half-bridge

CRBasic Example 60. PT100 in Three‐wire Half‐bridge 

'See FIGURE. PT100 in Three-Wire Half-Bridge

(p. 260)

for wiring diagram.

Public

Rs_Ro

Public

Deg_C

BeginProg

Scan

(1,Sec,0,0)

'BrHalf3W(Dest,Reps,Range1,SEChan,ExChan,MPE,Ex_mV,True,0,250,100.93,0)

BrHalf3W

(Rs_Ro,1,mV25,1,Vx1,1,2200,True,0,250,100.93,0)

'PRTCalc(Destination,Reps,Source,PRTType,Mult,Offset)

PRTCalc

(Deg_C,1,Rs_Ro,1,1.0,0)

NextScan

EndProg

 

7.8.18.2.4 PT100 in Four-Wire Full-Bridge

Example Shows:

• How to measure a PRT in a four-wire full-bridge

Advantages:

• Uses half as many input channels as four-wire half-bridge.

Example PRT Specifications:

• Alpha = 0.00392 (PRTType 2)

This example measures a 100 ohm PRT in a four-wire full-bridge, as shown in
figure PT100 in Four-Wire Full-Bridge

(p. 263),

using CRBasic instruction

BRFull(). In this example, the PRT is in a constant-temperature bath and the
measurement is to be used as the input for a control algorithm.

As described in table Resistive Bridge Measurements with Voltage Excitation (

p.

296

), the result of BRFull() is X,

X = 1000 V

S

/V

X

 

Advertising