6 offset voltage compensation, 1 input and excitation reversal – Campbell Scientific CR1000 Measurement and Control System User Manual

Page 282

Advertising
background image

Section 8. Operation

282 

 

 

8.1.2.6 Offset Voltage Compensation

Analog measurement circuitry in the CR1000 may introduce a small offset
voltage to a measurement. Depending on the magnitude of the signal, this offset
voltage may introduce significant error. For example, an offset of 3 μV on a 2500-
mV signal introduces an error of only 0.00012%; however, the same offset on a
0.25-mV signal introduces an error of 1.2%.

The primary source of offset voltage is the Seebeck effect, which arises at the
junctions of differing metals in electronic circuits. Secondary sources of offset
voltages are return currents incident to powering external devices through the
CR1000. Return currents create voltage drop at the ground terminals that may be
used as signal references.

CR1000 measurement instructions incorporate techniques to cancel these
unwanted offsets. The table Analog Measurement Instructions and Offset Voltage
Compensation Options

(p. 282)

lists available options.

Table 53. Analog Measurements and Offset Voltage Compensation


CRBasic

Voltage

Measurement

Instruction


Input

Reversal

(

RevDiff

=

True

)


Excitation

Reversal

(

RevEx

=

True

)

Measure

Ground

Reference

Offset

(

MeasOff

=

True

)

Background

Calibration

(RevDiff = False)

(RevEx = False)

(MeasOff = False)

VoltDiff()

*

*

VoltSe()

* *

TCDiff()

*

*

TCSe()

* *

BrHalf()

* *

BrHalf3W()

* *

Therm107()

* *

Therm108()

* *

Therm109()

* *

BrHalf4W()

* * *

BrFull()

* * *

BrFull6W()

* * *

AM25T()

* * *

 

8.1.2.6.1 Input and Excitation Reversal

Reversing inputs (differential measurements) or reversing polarity of excitation
voltage (bridge measurements) cancels stray voltage offsets. For example, if there
is a 3-

μV offset in the measurement circuitry, a 5-mV signal is measured as 5.003

mV. When the input or excitation is reversed, the measurement is -4.997 mV.
Subtracting the second measurement from the first and dividing by two cancels
the offset:

5.003 mV - (-4.997 mV) = 10.000 mV
10.000 mV / 2 = 5.000 mV.

Advertising