Section 8. Operation

279

 

where

Gain Error = ± (2500 * 0.0006) 

= ±1.5 mV 

and

Offset Error = 1.5 • 667 µV + 1 µV = 1.00 mV 

Therefore,

Error = Gain Error + Offset Error  

= ±1.5 mV + 1.00 µV 

= ±2.50 mV 

In contrast, the error for a 500‐mV input under the same constraints is 
±1.30 mV.  The figure Voltage Measurement Accuracy

 (p. 279

illustrates 

the total error with respect to voltage measurements for the ±2500‐mV 
range. 

Figure 87: Voltage measurement accuracy (0° to 40°C)