Section 7. Installation

164 

 

 

CRBasic Example 31. FieldCalStrain() Calibration Demonstration 

'Program to measure quarter bridge strain gage

'Measurements

Public

Raw_mVperV

Public

MicroStrain

'Variables that are arguments in the Zero Function

Public

Zero_Mode

Public

Zero_mVperV

'Variables that are arguments in the Shunt Function

Public

Shunt_Mode

Public

KnownRes

Public

GF_Adj

Public

GF_Raw

'----------------------------- Tables ----------------------------

DataTable

(CalHist,NewFieldCal,50)

SampleFieldCal

EndTable

'//////////////////////////// PROGRAM ////////////////////////////

BeginProg

'Set Gage Factors

GF_Raw = 2.1
GF_Adj = GF_Raw

'The adj Gage factors are used in the calculation of uStrain

'If a calibration has been done, the following will load the zero or

'Adjusted GF from the Calibration file

LoadFieldCal

(True)

Scan

(100,mSec,100,0)

'Measure Bridge Resistance

BrFull

(Raw_mVperV,1,mV25,1,Vx1,1,2500,True ,True ,0,250,1.0,0)

'Calculate Strain for 1/4 Bridge (1 Active Element)

StrainCalc

(microStrain,1,Raw_mVperV,Zero_mVperV,1,GF_Adj,0)

'Steps (1) & (3): Zero Calibration

'Balance bridge and set Zero_Mode = 1 in numeric monitor. Repeat after

'shunt calibration.

FieldCalStrain

(10,Raw_mVperV,1,0,Zero_mVperV,Zero_Mode,0,1,10,0 ,microStrain)

'Step (2) Shunt Calibration

'After zero calibration, and with bridge balanced (zeroed), set

'KnownRes = to gage resistance (resistance of gage at rest), then set

'Shunt_Mode = 1. When Shunt_Mode increments to 3, position shunt resistor

'and set KnownRes = shunt resistance, then set Shunt_Mode = 4.

FieldCalStrain

(13,MicroStrain,1,GF_Adj,0,Shunt_Mode,KnownRes,1,10,GF_Raw,0)

CallTable

CalHist

NextScan

EndProg

 