Campbell Scientific CR1000 Measurement and Control System User Manual

Page 308

Advertising
background image

Section 8. Operation

308 

 

 

Error Calculations with Input Reversal = True

µV error = gain term + offset term 

= (830.7 µV * 0.12%) + (1.5 * 0.67 µV + 1.0 µV) 

= 0.997 µV + 2.01 µV 

= 3.01 µV (= 0.071 °C) 

Error Calculations with Input Reversal = False

µV Error = gain term + offset term 

= (830.7 µV * 0.12%) + (3 * 0.67 µV + 2.0 µV) 

= 0.997 µV + 4.01 µV 

= 5.01 µV (= 0.12 °C) 

Input Error Examples: Type K Thermocouple @ 1300°C

Error in the temperature due to inaccuracy in the measurement of the
thermocouple voltage increases at temperature extremes, particularly when the
temperature and thermocouple type require using the ±200|250 mV range. For
example, assume type K (chromel-alumel) thermocouples are used to measure
temperatures around 1300°C.

These examples demonstrate that at temperature extremes, input offset error is
much less than input gain error because the use of a larger input range is required.

Conditions

CR1000 module temperature, ‐25 to 50°C 

Temperature = 1300°C 

Reference temperature = 25°C 

Delta T (temperature difference) = 1275°C 

Thermocouple output multiplier at 1300°C = 34.9 µV °C

‐1

 

Thermocouple output = 1275°C * 34.9 µV °C

‐1

 = 44500 µV 

Input range = ±250 mV 

Error Calculations with Input Reversal = True

µV error = gain term + offset term 

= (44500 µV * 0.12%) + (1.5 * 66.7 µV + 1.0 µV) 

= 53.4 µV + 101.0 µV 

= 154 µV (= 4.41 °C) 

Advertising