Abort status retention, Run completion on crystal failure, Powerful system interface – INFICON MDC-360 Thin Film Deposition Controller User Manual

Page 12: Power supply noise tolerance, International standard power connector, Field upgradable, Abort status retention -2, Run completion on crystal failure -2, Powerful system interface -2, Power supply noise tolerance -2

Advertising
background image

MDC-360 DEPOSITION CONTROLLER

1.2.7

ABORT STATUS RETENTION

In the event that the MDC-360 does abort during the deposition process, pertinent
information is stored at the time of abort. The process can be resumed after the
problem is corrected.

1.2.8

RUN COMPLETION ON CRYSTAL FAILURE

The extensive monitoring and abort functions are designed to protect the system
and/or process from serious and hopefully infrequent malfunctions of the
deposition system. A condition which need not cause an abort is the condition of
crystal failure. The MDC-360 can be set to abort upon crystal failure, or run to
completion using a backup crystal or time/power method.

1.2.9

POWERFUL SYSTEM INTERFACE

Fully programmable discrete inputs and outputs permit the MDC-360 to be easily
interfaced into deposition systems controlling the most complex processes. Also,
source control outputs are fully isolated avoiding ground loop problems. The
MDC-360 also supports input from an optical monitor for optical termination of
film thickness.

1.2.10 POWER SUPPLY NOISE TOLERANCE

Integral RFI filter and large energy storage capacitors will tolerate high levels of
power supply noise and power interruptions of 700 ms or less without effect.

1.2.11 INTERNATIONAL STANDARD POWER CONNECTOR

The power connector is internationally approved and meets IEC (International
Electrotechnical Commission) standards. It allows selection of input power
voltages ranging from 100 to 240 volts at a frequency of 50 or 60 Hz and includes
an integral RFI filter.

1.2.12 FIELD

UPGRADABLE

Plug-in interface boards and option boards allow the basic unit to be upgraded in
the field to the maximum system level.

GENERAL DESCRIPTION

1-2