MDC-360 DEPOSITION CONTROLLER

4-41

me

(minutes)

Data Points/Minute

Aprox. Storage Ti

30 900
60 450

120 225
300 90
600 45

8. Time (00:00-23:59)

This parameter sets the system time. Time is entered in 24-hour format without a
digit separator “:”. For example, to enter 1:05 PM you must enter “1305”. This
parameter is only visible when the data logging option is installed.

9. Date (01/01/00-12/31/99)

This parameter se

ts the system date in month/day/year format. The complete date

is parameter is only visible when the data logging option is installed.

must be entered without the digit separator “/” character, and with two digits for
each of the month, day and year. For example, to enter 5/2/94, you must enter
“050294”. Th

PROGRAMMING AND CONTROLLER SETUP