Carbon monoxide transmitter, Overview, Specifications – RKI Instruments Pioneer 4W User Manual

Page 104

Advertising
background image

Pioneer Operator’s Manual

CO Transmitter • 1

Carbon Monoxide Transmitter

Overview

This detection insert describes the carbon monoxide (CO) transmitter. This insert also
describes how to install, start up, maintain, and calibrate the transmitter and run the
Channel Setup program. A parts list at the end of this insert lists replacement parts and
accessories for the CO transmitter.

Specifications

Table 1 lists specifications for the CO transmitter. See the specifications in Chapter 1,
Introduction, for specifications that apply to the Pioneer Gas Monitor.

Table 1: Specifications

Target Gas

Carbon monoxide (CO)

Area Classification

Explosionproof for Class I, Groups B, C, and D

Sampling Method

Diffusion

Signal Output

4 to 20 mA

Detection Range

0 to 300 PPM (parts per million)*

Alarm 1 Setpoint

25 PPM*

Alarm 2 Setpoint

50 PPM*

Alarm 3 Setpoint

100 PPM*

Response Time

90% in 30 seconds

*

These specifications represent RKI standard settings; however, you can adjust them in the Channel Setup program.