Pioneer Gas Monitor Operator’s Manual

Description • 5

Fuses

The Pioneer includes up to two AC fuses and one DC fuse.

AC fuse(s)

Two AC fuse holders are above the power switch (near the upper left corner of the main
circuit board). The AC fuse(s) protects the AC line circuitry from short circuit or overload.

For a 115 VAC power source, a 1-amp fuse is installed in the top fuse holder. For a 220
VAC power source, two 1/2-amp fuses are installed in the top and bottom fuse holders.

NOTE:

The

standard

configuration is for 115 VAC operation. 220 VAC operation is an

optional

configuration and must be specified when ordering.

DC fuse

The DC fuse is to the right of the fourth analyzer card socket (near the lower right corner
of the main circuit board). The DC fuse protects the DC line circuitry from short circuit or
overload. The DC fuse is rated at 3 amps.

Power Switch

The power switch is directly below the AC fuse. The power switch turns the incoming
power source on and off at the Pioneer. When the switch is up, the power switch is on. A
round polycarbonate guard is mounted to the switch bushing to protect against accidental
contact with live parts on the circuit board near the switch.

Transformer

The transformer is to the right of the AC fuse (under the display screen). The transformer
receives the incoming AC voltage and converts it to a nominal DC voltage.

Ground Stud

The ground stud extends from the bottom left corner of the main circuit board. Remove
the nut to install a ground wire terminated with a #10 lug, then replace the nut.

CAUTION:

Do not use the

GND

terminal to connect the ground wire. Use the ground stud

described above.

Terminal Strips

The Pioneer includes four terminal strips for external wiring connections. See Chapter 3,
Installation and Start Up, for wiring procedures.

NOTE:

The analyzer card also includes three terminal strips. They are described at the

end of this chapter.

AC terminal strip

The 3-point AC terminal strip is below and to the right of the power switch. You connect
the incoming AC power source to the AC terminal strip. The AC terminal strip is labeled

TB1

on the main circuit board. A polycarbonate terminal cover protects against

accidental contact with the live terminal screws once power to the unit is turned on.