Pioneer Operator’s Manual (PRELIMINARY)

Oxygen Transmitter • 11

Channel status screen

This screen indicates the status of Channel 1. For installed channels the default setting is

ACTIVE/ONLINE

. For empty channels, the default setting is

NO SENSOR/AMP

. The

third setting,

SILENCED/OFFLINE

, displays the gas reading as XXXX; and the buzzer,

status lights, and relays are disabled for the channel.

To accept the channel status setting and continue the program, press the ENTER
button.

The program displays the Fullscale Reading, Unit of Measure, and Target Gas screen.

To update the channel status setting:

1.

Use the UP/YES and DOWN/NO buttons to change the setting.

2.

Press the ENTER button to accept the new setting and continue the program.

The program displays the Fullscale Reading, Unit of Measure, and Target Gas screen.

NOTE:

If you select a channel status of

NO SENSOR/AMP

, the program continues with

the next channel.

Fullscale reading, unit of measure, and target gas screen

This screen indicates the fullscale reading, unit of measure, and target gas for Channel 1.
The default settings for the oxygen channel are

25.0

(fullscale reading),

%VOL

(unit of

measure), and

OXY

(target gas).

To accept the settings and continue the program, press the ENTER button three times.

The program displays the Channel Type screen.

To update the target gas, unit of measure, and/or fullscale reading setting:

1.

Use the ENTER and ESCAPE buttons to select the setting you want to change.
(The setting flashes when it is selected.)

2.

Use the UP/YES and DOWN/NO buttons to change the setting.

3.

Press the ENTER button until the program displays the Channel Type screen.