20 • Operation

Pioneer Gas Monitor Operator’s Manual

NOTE:

You can select normally energized or normally de-energized relays in the
Channel Setup program. The following sections describe the default setting for
the relays: normally de-energized. The fail relay is factory-set for normally

energized and is not user-selectable.

Alarm 1 Condition

This section describes the audible and visual indications for an alarm 1 condition and
suggests response to an alarm 1 condition.

Alarm 1 condition indications

When the gas reading of an active channel reaches the alarm 1 setpoint, the Pioneer senses
an alarm 1 condition. The Pioneer alerts you to an alarm 1 condition as follows:

the ALARM 1 light turns on

the gas reading in alarm 1 condition flashes

the buzzer sounds a pulsing tone

the common alarm 1 relay energizes

the alarm 1 relay on the applicable analyzer card energizes

Responding to an alarm 1 condition

This section suggests response to an alarm 1 condition.

1.

Follow your established procedure for a low level combustible or toxic gas condition
or a decreasing oxygen content condition.

Table 4-1: Visual and Audible Alarm Indications

Condition

Cause

Visual Indication

Audible

Indication

Alarm 1

Increasing (decreasing for O

2

) gas reading

at or above the alarm 1 setpoint

• ALARM 1 light is on

• Gas reading flashes

Pulsing tone

Alarm 2

Increasing (decreasing for O

2

) gas reading

at or above the alarm 2 setpoint

• ALARM 1 and ALARM 2

lights are on

• Gas reading flashes

Faster

pulsing tone

than Alarm 1

Alarm 3

Increasing gas reading at or above the

alarm 3 setpoint

• ALARM 1, ALARM 2,

and ALARM 3 lights are
on*

• Gas reading flashes

Faster

pulsing tone

than Alarm 2

Fail

• Disconnected or misconnected

detector wiring

• Disconnected or misconnected

analyzer card

• Display reading below fail setpoint

• Defective components

• FAIL light is on

Steady tone

* The ALARM 1 and ALARM 2 lights are not on for oxygen channels.