Maintenance – RKI Instruments Pioneer 4W User Manual

Page 116

Advertising
background image

Pioneer Operator’s Manual

CO Transmitter • 13

Channel setup exit screen

This screen saves or cancels any changes you made in the Channel Setup program.

To save changes

and exit the program, press the UP/YES or DOWN/NO button.

The program displays a confirmation message, then the normal screen appears. The
Pioneer is in normal operation.

To cancel changes

and exit the program, press the ESCAPE button.

The program displays a confirmation message, then the normal screen appears. The
Pioneer is in normal operation.

To setup another channel

,

press the ENTER button.

The program asks if you want to setup the next channel. Press the ENTER button to
continue.

Maintenance

This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.

Preventive Maintenance

This section describes a preventive maintenance schedule to ensure the optimum
performance of the CO transmitter. It includes daily, monthly, and quarterly procedures.

Daily

1.

Verify that the pilot light on the applicable analyzer card is flashing. If the pilot light is
not flashing, see the following troubleshooting guide in this insert.

NOTE:

If the Pioneer’s main PILOT light is not on, see the Troubleshooting guide in

Chapter 5, Maintenance.

2.

Verify a display reading of 0 PPM. Investigate significant changes in the display
reading.

Monthly

This procedure describes a test to verify that the CO transmitter responds properly to
carbon monoxide.

Preparing for the response test

1.

Verify that the display reading(s) for the channel(s) you plan to test is 0.

If the display reading(s) is not zero, set the zero reading of the applicable
transmitter(s) as described in the Start Up section of this insert, then continue this
procedure.

E S C A P E

=

A b o r t

S e t

E N T E R =

S a v e

&

E x i t

=

u p

U P / D O W N

N e x t

Advertising