RKI Instruments Pioneer 4W User Manual
Page 98

Oxygen Transmitter • 22
Pioneer Operator’s Manual (PRELIMINARY)
10. Reconnect the wiring from the Pioneer to the interconnect terminal strip as shown in
Table 3 and Figure 3, Wiring the Oxygen Transmitter to the Pioneer Gas Monitor.
11. Reconnect the wiring from the (2-point) sensor terminal block in the detector junction
box to the amplifier’s detector terminal strip as shown in Table 4 and Figure 3, Wiring
the Oxygen Transmitter to the Pioneer Gas Monitor.
12. Place the Pioneer’s power switch in the ON position.
13. Calibrate the oxygen transmitter as described in the Calibration section of this insert.
Table 3: Reconnecting the Oxygen Amplifier to the Pioneer Gas Monitor
Amplifier
Interconnect Terminal Strip
Pioneer Analyzer Card
4/20 FB
4-20 (Terminal 5)
24V +
24 V (Terminal 6)
Table 4: Reconnecting the Oxygen Detector to the Amplifier
Sensor
Terminal Block
Amplifier
Interconnect Terminal Strip
+ (white)
OXY W
— (green)
OXY G