Alarm 2 condition indications, Responding to an alarm 2 condition – RKI Instruments Pioneer 4W User Manual

Page 25

Advertising
background image

Pioneer Gas Monitor Operator’s Manual

Operation • 21

2.

After the gas reading falls below the alarm 1 setpoint, press the reset switch to reset
the alarm 1 circuit. Resetting the alarm 1 circuit silences the buzzer, turns off the
ALARM 1 light, and de-energizes the common and analyzer card alarm 1 relays.

NOTE:

To silence the buzzer while in an alarm 1 condition, press the reset switch.
You cannot de-energize the alarm 1 relays until the gas reading falls below

(above for oxygen) the alarm 1 setpoint.

Alarm 2 Condition

This section describes the audible and visual indications for an alarm 2 condition and
suggests response to an alarm 2 condition.

Alarm 2 condition indications

When the gas reading of an active channel reaches the alarm 2 setpoint, the Pioneer senses
an alarm 2 condition. The Pioneer alerts you to an alarm 2 condition as follows:

the ALARM 2 light turns on

the gas reading in alarm 2 condition continues to flash

the buzzer sounds a faster pulsing tone than an alarm 1 condition

the common alarm 2 relay energizes

the alarm 2 relay on the applicable analyzer card energizes

Responding to an alarm 2 condition

This section suggests response to an alarm 2 condition.

1.

Follow your established procedure for a mid level combustible or toxic gas condition
or a severe decreasing oxygen content condition.

2.

After the gas reading falls below the alarm 2 setpoint, press the reset switch to reset
the alarm circuit. Resetting the alarm circuit turns off the ALARM 2 light, and
de-energizes the common and analyzer card alarm 2 relays.

NOTE:

To silence the buzzer while in an alarm 2 condition, press the reset switch.
You cannot de-energize the alarm 2 relays until the gas reading falls below

(above for oxygen) the alarm 2 setpoint.

Alarm 3 Condition

This section describes the audible and visual indications for an alarm 3 condition and
suggests response to an alarm 3 condition.

Alarm 3 condition indications

When the gas reading of an active channel reaches the alarm 3 setpoint, the Pioneer senses
an alarm 3 condition. The Pioneer alerts you to an alarm 3 condition as follows:

the ALARM 3 light turns on

the gas reading in alarm 3 condition continues to flash (the gas reading for O

2

channels begins flashing)

the buzzer sounds a faster pulsing tone than an alarm 1 or alarm 2 condition

the common alarm 3 relay energizes

the alarm 3 relay on the applicable analyzer card energizes