32 • Operation

Pioneer Gas Monitor Operator’s Manual

To display alarm sets:

The Display Setpoints and Readings program displays the alarm sets for all active
channels. The alarm sets include the alarm logic, alarm silence, alarm setpoint, and alarm
type settings.

NOTE:

You can change the alarm logic and alarm silence settings in the Instrument
Setup program. You can change the alarm setpoint and alarm type settings in the

Channel Setup program.

1.

Press and hold down the reset switch to enter the program.

2.

When the ALARM SETS screen appears on the display screen, release the reset switch.

The alarm logic setting (LATCHING or AUTO-RESET) displays for all active
channels.

3.

Press and release the reset switch to display the alarm silence setting (ON or OFF) for
all active channels.

4.

Continue pressing and releasing the reset switch to display the setpoint and alarm
type for the fail, alarm 1, alarm 2, and alarm 3 alarms of channel 1.

5.

Repeat step 4 to display the setpoint and alarm type for the remaining active channels.

6.

Press and release the reset switch to return to the normal screen.

To perform the self check:

The Pioneer includes a self check feature that makes sure the analyzer card(s) and
microprocessor are operating correctly.

1.

Press and hold down the reset switch to enter the program.

2.

When the SELF CHECK screen appears on the display screen, release the reset switch.
The Pioneer performs the self check procedure and returns to the normal screen. This
process takes approximately 5 seconds.

To reset Minimum/Maximum readings:

If you reset the minimum and maximum readings, you cannot retrieve the minimum and
maximum readings previously stored by the Pioneer.

1.

Press and hold down the reset switch to enter the program.

2.

When the RESET MIN/MAX screen appears on the display screen, release the reset
switch.

3.

The message “MIN/MAX RESET DONE” appears on the display screen, then the
program returns to the normal screen.