Running the instrument setup program, Entering the instrument setup program, Alarm on delay screen – RKI Instruments Pioneer 4W User Manual

Page 28: Press the enter button to continue the program, Alarm off delay screen, The buzzer is not working properly

Advertising
background image

24 • Operation

Pioneer Gas Monitor Operator’s Manual

Running the Instrument Setup Program

This section describes how to enter the program, display and change program settings,
and save or cancel changes to the program settings.

Entering the instrument setup program

1.

Open the housing door, and locate the program buttons to the right of the display
screen.

2.

Press the ENTER and ESCAPE buttons simultaneously three times.

The display screen indicates that you are in the Instrument Setup program.

3.

Press the ENTER, UP/YES, or DOWN/NO button to continue.

The program displays the Alarm ON Delay screen.

Alarm on delay screen

This screen indicates the length of time that the Pioneer delays alarm indications (buzzer
and relays) when it senses an alarm 1, alarm 2, or alarm 3 condition. You can choose from
a range of 0 Seconds to 60 Minutes. The default setting is

0 Minutes, 1 Seconds

.

To accept the alarm on delay setting and continue the program, press the ENTER
button.

The program displays the Alarm OFF Delay screen.

To update the alarm on delay setting:

1.

Use the UP/YES and DOWN/NO buttons to display the setting you want.

2.

Press the ENTER button to continue the program.

The program displays the Alarm OFF Delay screen.

Alarm off delay screen

This screen indicates the length of time that the Pioneer delays turning off alarm
indications (buzzer and relays) after an alarm 1, alarm 2, or alarm 3 condition passes. You
can choose from a range of 0 Seconds to 60 Minutes. The default setting is

0 Minutes,

0 Seconds

.

To accept the alarm off delay setting and continue the program, press the ENTER
button.

The program displays the Relay Action screen.

To update the alarm off delay setting:

1.

Use the UP/YES and DOWN/NO buttons to display the setting you want.

2.

Press the ENTER button to continue the program.

The program displays the Relay Action screen.

A l a r m

O N

D e l a y

:

0

M i n u t e s

1

S e c o n d s

A l a r m

O F

D e l a y :

0

M i n u t e s

0

S e c o n d s

F