Pioneer Gas Monitor Operator’s Manual

Operation • 27

Calibration time out screen

The Pioneer’s Calibrate program includes two separate time-out features. The first
time-out feature is a general 5-minute time-out. The Pioneer automatically returns to the
normal screen if you do not move from one Calibrate screen to another within 5 minutes.
You cannot adjust this setting, and it affects all Calibrate screens except the Fresh Air and
Calibration screens.

The second time-out feature affects the Fresh Air and Calibration screens only. These are
the screens you use during calibration to adjust the zero and span readings (either at the
transmitter or the Pioneer). The Pioneer automatically returns to the normal screen if you
do not move from the Fresh Air or Calibration screen within the time-out setting specified
in the screen below. You can choose from a range of 10 to 100 minutes (in increments of
10). The default setting is

10 Minutes

.

To accept the calibration time out setting and continue the program, press the ENTER
button.

The program displays the Instrument ID screen.

To update the calibration time out setting:

1.

Use the UP/YES or DOWN/NO button to display the setting you want.

2.

Press the ENTER button to continue the program.

The program displays the Instrument ID screen.

Instrument ID screen

This screen indicates the ID number (address) for this Pioneer. The instrument ID is
recorded in the datalog when the datalog records instrument readings. This setting is
useful if you have more than one Pioneer networked together. You can choose from a
range of 1 to 255. The default setting is

1

.

To accept the instrument ID setting and continue the program, press the ENTER
button.

The program displays the Noise Filter screen.

To update the instrument ID setting:

1.

Use the UP/YES or DOWN/NO button to display the setting you want.

2.

Press the ENTER button to continue the program.

The program displays the Noise Filter screen.

:

C a l i b r a t i o n

M e n u

T i m e - O u t

1 0

M i n u t e s

I N S T R U M E N T

I D

F O R

R S 2 3 2

&

R S 4 8 5 :

1