Detaching and attaching pump unit – RKI Instruments GD-70D Series User Manual

Page 13

Advertising
background image

1 Overview

1-5. Units description

- 13 -

GD-70D

Semiconductor

type (SGU)








• Before using the sensor unit, it needs to be warmed up for a specified time.

The sensor unit is warmed up sufficiently in our factory before it is delivered to
you. Therefore, after you receive the sensor unit, please use it as soon as
possible so that unpowered time is minimized.

• The warm-up (powered) time before using the sensor unit is related to the

unpowered time.

Suggested powered time

Unpowered time

SGU-8541 (H2)

SGU (Others)

10 minutes or less

10 minutes or more

2 hours or more

1 hour or less

30 minutes or more

2 hours or more

24 hours or less

1 hour or more

4 hours or more

72 hours or less

4 hours or more

24 hours or more

10 days or less

2 days or more

2 days or more

Less than 1 month

7 days or more

7 days or more

Less than 3 months

14 days or more

14 days or more

3 months or more

1 month or more

1 month or more

• To store the sensor unit under unpowered state, it must be stored under

normal temperature/humidity in a clean place away from direct sunlight.

Galvanic cell type

(OSU)





• Do not disassemble the sensor unit because it contains electrolyte. If contact

occurs, rinse the area immediately with a large quantity of water.

• The sensor unit must be stored under normal temperature/humidity in a clean

place away from direct sunlight.



<Detaching and Attaching Pump Unit>


After detaching the sensor unit, push down to open the pump stopper. Pinch the part labeled "PULL" at the
top of the pump unit, and pull it out.








When attaching the pump unit to the main unit, position the
grooves of the pump unit at the rails of the main unit and push the
pump unit.
Do not push the center diaphragm. Push the part labeled "PULL".
And as the connecting point (flow path) of the pump unit is
greased, be sure to check for any dust.









connection

(flow path)

×

Do not push here.

"PULL"

groove

rail