7 Regular Maintenance Mode

GD-70

- 48 -





7

Regular Maintenance Mode












Mode Item

LCD

Display

Details

Gas Introduction Display

2-0 GAS TEST

Perform the gas introduction test in the regular
maintenance mode.

Zero Adjustment

P65

2-1 ZERO

Perform the zero adjustment.

Span Adjustment

P65

2-2 SPAN

Perform the span adjustment.

Last Calibrated Date

2-3 LAST CAL

Show the last calibrated date.

Bias Voltage (Element
Voltage)

2-4 BIAS
(2-4 E VOLT)

Show the bias voltage.
(Show the element voltage.)

Flow Rate Setting
(adjusted to 0.5 L/min)

P67

2-5 DEF
FLOW

Set the flow sensor with the flow rate at 0.5 L/min.

Pump Ratio/Flow Rate
Indicator

P67

2-6 FLOW

Show the output and flow rate of the current pump.

Detector Temperature

2-7 TEMP

Show the current temperature of the installation
environment.

Suggested Warm-up
Completion Date/Time

2-8
WARMTIME

Show the suggested warm-up completion for
semiconductor type (SGU).

Maintenance
Mode
(regular
maintenance
mode)

Environmental Setting 1

P53

2-9 SETTING1

Operation setting
• INHIBIT setting (INHIBIT)
• Alarm value setting (ALM P)

P54

• Alarm delay time setting (ALM DLY)

• Regular replacement operation (pump stop)

(MAINTE)
• Fault alarm test (F TEST)

P54

WARNING

After the adjustment is completed, do not forget to press the MODE key to return to the detection mode.
(If the detector remains in the maintenance mode, it automatically returns to the detection mode in ten
hours.)