0ppm, Warm up, Sih4 – RKI Instruments GD-70D Series User Manual

Page 31

Advertising
background image

3 How to Operate

3-2. How to start the detector

- 31 -

GD-70D

<Start-up Procedures (approximately 25 seconds for system
check of the detector and alarm deactivation)>

Power On

PW

A1

A2

F

LCD

Initial Clear

● ○ ○ ○

──────

WARM UP

Gas
Specifications
Display

● ○ ○ ○

15.0ppm

SIH4

Detection Mode

● ○ ○ ○

0.0ppm

SIH4



































WARNING

• When Oxygen (OSU) is selected, approximately AIR

(normal oxygen concentration) (20.9 vol%) is output after
the detector is started in the atmosphere.
For a specification where an alarm is triggered by the AIR
(normal oxygen concentration) output (e.g., 0 - 5 vol%),
be careful of an alarm issue after the detector is started.

CAUTION

• Do not turn off the detector during the initial clear. The detector is reading the sensor memory during

the initial clear.

• If a new sensor unit is installed or the sensor unit is replaced after the detector is started, the sensor

unit must be warmed up for a specified period which is determined depending on the type of the
sensor unit.
When the semiconductor type sensor unit (SGU) is selected, the suggested warm-up completion
time is displayed in the maintenance mode.
During the warm-up, the alarm activation and output signals are unstable. Provide a prior notification
to the related sections so that they can prepare for false abnormalities.

• Because the pyrolyzer unit (PLU-70) must be warmed up for one hour, please warm it up with the

detector.

• After the warm-up is completed, verify that the reading on the flow rate indicator corresponds to the

specified flow rate, and then perform a gas calibration.

PW:POWER
A1:ALM1
A2:ALM2
F:FAULT

:Lamp on

:Lamp off