RKI Instruments GD-70D Series User Manual

Page 34

Advertising
background image

3 How to Operate

3-5. Modes

GD-70D

- 34 -

Mode Item

LCD

Display

Details

Regular

Maintenance

Mode Switching

1-8 M MODE

Switch to the regular maintenance mode.

Gas Introduction Display

2-0 GAS TEST

Perform the gas introduction test in the regular maintenance mode.

Zero Adjustment

2-1 ZERO

Perform the zero adjustment.

Span Adjustment

2-2 SPAN

Perform the span adjustment.

Last Calibrated Date

2-3 LAST CAL

Show the last calibrated date.

Bias Voltage (Element
Voltage)

2-4 BIAS
(2-4 E VOLT)

Show the bias voltage.
(Show the element voltage.)

Flow Rate Setting
(adjusted to 0.5 L/min)

2-5 DEF FLOW

Set the flow sensor with the flow rate at 0.5 L/min.

Pump Ratio/Flow Rate
Indicator

2-6 FLOW

Show the output and flow rate of the current pump.

Detector Temperature

2-7 TEMP

Show the current temperature of the installation environment.

Suggested Warm-up
Completion Date/Time

2-8 WARMTIME

Show the suggested warm-up completion for semiconductor type
(SGU).

Environmental Setting 1

2-9 SETTING1

Operation setting

INHIBIT setting (INHIBIT)

Alarm value setting (ALM P)

Alarm delay time setting (ALM DLY)

Regular replacement operation (pump stop) (MAINTE)

Fault test (F TEST)

Environmental Setting 2

2-10 SETTING2

Functions setting

Address setting (ADDRESS)

Date/Time setting (DAY TIME)

Zero suppression value setting (SUPPRESS)

Zero suppression system setting (SUP TYPE)

Alarm test time contact setting (TEST RLY)

Alarm test time external output setting (TEST4-20)

Energized/De-energized setting (RLY PTRN)

Alarm type setting (ALM TYP)

Alarm activation setting (ALM PTRN)

Alarm value limiter setting (AL LIMIT)

Fault activation setting (FLT PTRN)

Flow rate auto-adjustment setting (AT FLOW)

Zero follower ON/OFF setting (ZERO F)

24 hours zero follower ON/OFF setting (ZERO 24F)

Sensitivity correction ON/OFF setting (S ASSIST)

External output in maintenance mode setting (MNT OUT)

External output adjustment (MA 4-20)

Backlight setting (BK LIGHT)

ETHERNET setting (ETHERNET)

Pump drive level diagnosis ON/OFF setting (PUMP CK)

Pyrolyzer Data Display

2-11 PL DATA

When the pyrolyzer unit (PLU-70) is used, a variety of pyrolyzer data
is displayed. (See the operating manual for PLU-70)

Fault Investigation

2-12 FAULT

Not used

Maintenance
Mode
(regular
maintenance
mode)

Factory Mode Switching

2-13 F MODE

Not used