Zero follower function, Sensitivity correction function – RKI Instruments GD-70D Series User Manual

Page 41

Advertising
background image

4 Detection Mode

4-4. Other functions

- 41 -

GD-70D

<Zero Follower Function>


Some types of sensor used with the detector might
have sensitivity variations after being used for a long
period.
This function corrects the variation of the reading at
the zero point (zero drift) among the sensitivity
variations over time by a program manipulation to
stabilize the zero point, and works on the
electrochemical type (ESU) and pyrolysis-particle
type (SSU).
























<Sensitivity Correction
Function>


Some types of sensor used with the detector might
have sensitivity variations after being used for a long
period.
This function compensates the degradation of the
gas sensitivity among the sensitivity variations over
time. It works on the electrochemical type (ESU) and
makes the span adjustment by a program
manipulation based on the principled degradation
pattern.











* The zero adjustment is made periodically.

Zero Follower On

Zero Follower Off

* The zero drift of the sensor appears as the reading.

Reading

Zero point

0

Time

Zero point

0

Reading

Time

Sensitivity Correction Off

* The sensor sensitivity becomes degraded.

Sensitivity Correction On

* The span is lifted up periodically.

0

Time

Gas correction

Reading

Gas correction

0

Reading

Time