5. modes, Detection mode, Warning – RKI Instruments GD-70D Series User Manual

Page 33: Caution, Flow rate indicator

Advertising
background image

3 How to Operate

3-5. Modes

- 33 -

GD-70D








<Detection mode>

Flow Rate Indicator

Because the suction flow rate of the detector is automatically adjusted by the flow rate control function, the
flow rate, in principal, does not need to be controlled. As shown on the figure below, when the flow rate does
not correspond to the specified flow rate for some reasons, it is adjusted automatically.























3-5. Modes


Details on each mode are provided as follows. (* Operations are slightly different depending on the detector
type or sensor unit.)

Mode Item

LCD

Display

Details

Detection Mode -

Gas concentration
Gas name

Normal state

Gas Alarm
Test Mode

-

Gas concentration

Perform the alarm test.

Zero Adjustment (Span
Adjustment)

1-1 ZERO
(1-1 SPAN)

Perform the zero adjustment.
(In case of oxygen 0 - 25 vol%, perform the span adjustment.)

Setting Display

1-2 CONFIRM

Show the setting of typical menus.

First Alarm Setpoint (AL1)

Second Alarm Setpoint (AL2)

Alarm delay time

Zero suppression value

Zero Follower ON/OFF

Sensitivity Correction ON/OFF

Flow Rate Indicator

1-3 FLOW

Show the current flow rate.

Address Display

1-4 ADDRESS

Show the address.

Detector Version Display

1-5 70D VER

Show the program version of the main unit.

Unit Version Display

1-6 UNIT VER

Show the program version of the sensor unit.

Maintenance
Mode (user
mode)

Net Version Display

1-7 NET VER

Show the program version of the communication function.

WARNING

• When the detector enters each mode from the detection mode while an alarm is activated, the alarm

contact is released.

CAUTION

• If the automatic flow rate adjustment does not work (due to clogged tube or leak), messages such as

"FLOW" for an unstable flow rate or "E-05" for flow rate abnormalities are displayed. In this case, you
must identify the causes and take appropriate actions.

Flow rate

Flow rate

Normal