7 Regular Maintenance Mode

- 51 -

GD-70D

↓ ↑

● ○ ○

2- 5

DEF FLOW

MAINTENANCE

SET

1000

DEF FLOW

MAINTENANCE

2-5.DEF FLOW

Set the flow rate to the default
value when the flow sensor is
replaced, etc.
Normally, this is not used
because the detector has
already been adjusted.
On the other hand, if the flow
meter is set at other than the
specified flow rate, gases
cannot be detected properly
due to an incorrect flow rate.
(Press the MODE key to
cancel this menu.)

| ↑

| |

| |

↓ |

● ○ ○

2- 6

FLOW

MAINTENANCE

SET

50 %

0.50 L/M

MAINTENANCE

2-6.FLOW

Show the output and flow rate
of the current pump.

↓ ↑

● ○ ○

2- 7

TEMP

MAINTENANCE

SET

25.0

°C

TEMP

MAINTENANCE

2-7.TEMP

Show the temperature of the
detector.

↓ ↑

● ○ ○

2- 8

WARMTIME

MAINTENANCE

SET

12:00

2009.01.01

MAINTENANCE

2-8. WARMTIME

Show the suggested warm-up
completion date/time for
semiconductor type (SGU),
which requires a long
warm-up time.

↓ ↑

● ○ ○

2- 9

SETTING1

MAINTENANCE

SET

Environmental
Setting 1

⇒P53

2-9. SETTING1

Specify the environmental
setting 1.

↓ ↑

● ○ ○

2-10

SETTING2

MAINTENANCE

SET

Environmental
Setting 2

⇒P55

2-10.SETTING2

Specify the environmental
setting 2.

↓ ↑

● ○ ○

2-11

PL DATA

MAINTENANCE

SET

For information on
Pyrolyzer Data
Display, see
"PLU-70 Operating
Manual".

2-11.PL DATA

Show the pyrolyzer data when
the pyrolyzer unit (PLU-70) is
used.

↓ ↑