7 Regular Maintenance Mode

GD-70D

- 54 -

<Alarm Value Setting "2-9" - "SET 1">

Alarm Value Setting

PW

A1

A2

F

LCD

SET 1.ALM P

Press the SET key.

● ○ ○ ○

SET 1

ALM P

MAINTENANCE

First Alarm Value Setting

Select the number by pressing the ▲ or
▼ key, and then press the SET key to
confirm the value. (Press the MODE key
to skip this menu.)

● ○ ○ ○

5.0ppm

AL 1

MAINTENANCE

Second Alarm Value Setting

Select the number by pressing the ▲ or
▼ key, and then press the SET key to
confirm the value. (Press the MODE key
to skip this menu.)

● ○ ○ ○

10.0ppm

AL 2

MAINTENANCE

Return to

SET 1.ALM P


<Fault Alarm Test "2-9" - "SET 4">

PW

A1

A2

F

LCD

SET 4.F TEST

Press the SET key.

SET 4

F TEST

MAINTENANCE

OFF

F TEST

MAINTENANCE

↓ ↑

Fault Test ON/OFF

Select either ON/OFF.
Switch ON and press the SET key to
activate the fault alarm.
Return to OFF and press the SET key to
deactivate the fault alarm.
(You can also press the MODE key to
cancel this menu and go back to the
original state.)

ON

F TEST

MAINTENANCE








WARNING

Because the contact (fault) can be activated only by a fault alarm test in the maintenance mode, be careful
to perform the test. The fault alarm test cannot be performed during inhibit.