7 Regular Maintenance Mode

GD-70D

- 50 -

Regular Maintenance Mode

PW

A1

A2

F

LCD

User Mode

In "1-8.M MODE", press the
SET key.

● ○ ○

1- 8

M MODE

MAINTENANCE

Then press the SET key again
for three seconds.

● ○ ○

- - - -

M MODE

MAINTENANCE

Regular Maintenance
Mode

● ○ ○

2- 0

GAS TEST

MAINTENANCE

SET

0.2ppm

GAS TEST

MAINTENANCE

(Alternate Display)↑

0.2ppm

SIH4

MAINTENANCE

2-0.GAS TEST

Perform a test with the gas.
Similar to the detection mode,
the reading changes and the
alarm lamp lights up after the
gas is introduced, but the
contact is not activated.

| ↑

| |

↓ |

● ○ ○

2- 1

ZERO

MAINTENANCE

SET

Zero Adjustment

⇒P65

2-1.ZERO

Perform the zero adjustment.

↓ ↑

● ○ ○

2- 2

SPAN

MAINTENANCE

SET

Span Adjustment

⇒P66

2-2.SPAN

Perform the span adjustment.

↓ ↑

● ○ ○

2- 3

LAST CAL

MAINTENANCE

SET

12:00

2009.01.01

MAINTENANCE

2-3.LAST CAL

Show the last calibrated date.

↓ ↑

● ○ ○

2- 4

BIAS

MAINTENANCE

SET

0 mV

BIAS

MAINTENANCE

2-4.BIAS

Show the bias voltage "BIAS"
of electrochemical type (ESU)
or the element voltage
"EVOLT" of semiconductor
type (SGU).

↓ ↑

PW: POWER
A1: ALM1
A2: ALM2
F: FAULT

●: Lamp on
○: Lamp off