2 Installation

2-3. How to wire

GD-70D

- 24 -

Connection conditions

• Cable: 0.08 - 2.5mm

2

• Bare wire length: 8 - 9 mm

• Connecting tool: Dedicated handling lever (accessory)



















Compatible bar terminal

For a bar terminal, the following items are available.
• Bar terminal (ferrule): Model 216 Series (manufactured by WAGO)
• Crimping tool: Model VarioCrimp 4 (206-204) (manufactured by WAGO)






Dedicated

handling lever

CAUTION

• The specified bare wire length must be observed when the wire insulation is peeled off.

• Improper clamping of the wire due to a shorter bare wire length may cause defective electrification or

heating.

• Damaging the wire insulation due to a shorter bare wire length may cause defective electrification or

heating.

• Exposing the wire due to a longer bare wire length may cause defective insulation or a short circuit.

• Be careful not to break up the wire. If the wire is broken up when inserted to the terminal, this may

cause defective insulation or heating.

8 – 9 mm

CAUTION

• A bar terminal of the specified model must be used. Using other bar terminals invalidates the

guarantee of the performance.