8 Maintenance

8-2. Gas calibration method

GD-70D

- 64 -

8-2. Gas calibration method


Perform a gas calibration in each mode (zero adjustment mode
and span adjustment mode) using the calibration gas.

• Zero adjustment gas (collected in a gas sampling bag)

• Span adjustment gas (collected in a gas sampling bag)

• Gas sampling bag for exhaust




















<Zero Adjustment "2-1">

This is used to perform the zero adjustment.









NOTE

Before starting the zero adjustment, let the detector suck the zero adjustment gas and wait until the
indicator is stabilized.

WARNING

When the zero adjustment is performed in the atmosphere, check the atmosphere for freshness before
beginning the adjustment. If other gases exist, the adjustment cannot be performed properly, thus
leading to dangers when the gas leaks.

Exhaust
gas

GAS IN

GAS OUT

Calibration
gas

WARNING

After the adjustment is completed, do not forget to press the MODE key to return to the detection mode.