8 Maintenance

8-1. Maintenance intervals and items

GD-70

- 62 -





8

Maintenance





This is a critical instrument for disaster-proof and safety.
To maintain the performance of the detector and improve the reliability of disaster-proof and safety, perform
a regular maintenance.

NOTE

To use the pyrolyzer unit (PLU-70), also refer to the individual operating manual.


8-1. Maintenance intervals and items

• Daily maintenance: perform a maintenance before beginning to work.

• Monthly maintenance: perform a maintenance on the alarm circuit (alarm test) once a month.

• Regular maintenance: perform a maintenance once or more for every six months to maintain the

performance as a safety device.

Maintenance
item

Maintenance content

Daily

maintenance

Monthly

maintenance

Regular

maintenance

Power supply
check

Verify that the power lamp lights up.

○

○

○

Concentration
display check

Verify that the concentration display
value is zero (or 20.9 vol% on the oxygen
deficiency meter). When the reading is
incorrect, perform the zero adjustment
after ensuring that no other gases exist
around it.

○

○

○

Flow rate
check

See the flow rate indicator to check for
abnormalities.

○

○

○

Filter check

Check the dust filter for dust or clogging.

○

○

○

Alarm test

Inspect the alarm circuit by using the
alarm test function.

○

○

Span
adjustment

Perform the sensitivity calibration by
using the calibration gas.

○

Gas alarm
check

Check the gas alarm by using the
calibration gas.

○