8 Maintenance

8-3. Other adjustments/Cleaning method

- 67 -

GD-70D

8-3. Other adjustments/Cleaning method

<Flow Rate Manual Adjustment "2-6" and Flow Rate Default Set
"2-5">

The flow rate of the detector is automatically adjusted to 0.5 L/min. Turning off the auto-adjustment function
enables the manual adjustment. (See 2-10 - SET-11)
The manual flow rate adjustment can be performed in the regular maintenance mode "2-6.FLOW".

PW

A1

A2

F LCD

2-6.FLOW

Press the SET key.

2-6

FLOW

MAINTENANCE

50 %

0.50 L/M

MAINTENANCE

Flow Rate Adjustment

As the output is increased or decreased by
pressing the ▲ or ▼ key, the flow rate
changes accordingly. Adjust the flow rate to
0.5 L/min, and then press the SET key.
(Press the MODE key to cancel this menu.)


Regardless of the flow rate auto-adjustment or manual-adjustment, when the reading on the flow rate
indicator is incorrect (due to aging degradation, improper flow rate default set, or other reasons), it must be
set to provide the right flow rate.
To set the Flow Rate Indicator, prepare and connect a flow meter (which indicates 0.5 L/min precisely), and
perform the default set with the flow meter while it is indicating 0.5 L/min in the regular maintenance mode
"2-5.DEF FLOW".

PW

A1 A2 F

LCD

2-5.DEF FLOW

Press the SET key.

2- 5

DEF FLOW

MAINTENANCE

1000

DEF FLOW

MAINTENANCE

Flow Rate Adjustment/Default Set

Adjust the flow rate by pressing the ▲ or ▼
key until the connected flow meter indicates
0.5 L/min

± 10%, and then press the SET

key.
(Press the MODE key to cancel this menu.)












CAUTION

Perform the flow rate default set after verifying that the suction volume is 0.5 L/min by using a flow meter.

WARNING

After the adjustment is completed, do not forget to press the MODE key to return to the detection mode.