7 Regular Maintenance Mode

GD-70D

- 56 -

↓ ↑

● ○ ○

SET 5

TEST4-20

MAINTENANCE

SET

ON

TEST4-20

MAINTENANCE

↓↑

OFF

TEST4-20

MAINTENANCE

SET 5.TEST4-20

Set the external output for an
alarm test.
Select either ON/OFF, and
then press the SET key to
confirm the selection.
When ON is selected, the
external output is output even
during an alarm test.
When OFF is selected, the
output is the same as the
external output (selected in
SET 15) in the maintenance
mode.
* When 4 - 20 mA is selected
in SET 15, the output is made
even though OFF is selected.

| ↑

| |

| |

| |

↓ |

● ○ ○

SET 6

RLY PTRN

MAINTENANCE

SET

Energized/De-Ener
gized Contact
Setting

⇒P61

SET 6.RLY PTRN

Select either
energized/de-energized for
the contact.
(Energized/De-Energized
Contact Setting)

↓ ↑

● ○ ○

SET 7

ALM TYPE

MAINTENANCE

SET

L-LL

ALM TYPE

MAINTENANCE

↓↑

L-H

ALM TYPE

MAINTENANCE

↓↑

H-HH

ALM TYPE

MAINTENANCE

SET 7.ALM TYPE

Set the alarm type.
When the galvanic cell type
(OSU) is installed, you can
select either L-LL, L-H, or
H-HH.

| ↑

| |
| |
| |

| |

| |

↓ |

● ○ ○

SET 8

ALM PTRN

MAINTENANCE

↓ ↑

SET 8.ALM PTRN

This is a setting screen of the
gas alarm activation. Do not
change the setting when the
detector is used in a normal
way, because it determines
how the detector functions.
(Auto-reset setting: "nL")

● ○ ○

SET 9

AL LIMIT

MAINTENANCE

SET 9.AL LIMIT

This is a setting screen of the
alarm value limiter. Do not
change the setting when the
detector is used in a normal
way.
(ON setting)

↓ ↑