Set 15, Hold, 20ma – RKI Instruments GD-70D Series User Manual

Page 58: Set 16, Set 17, Save, Set 18, Mnt out, Ma 4-20, 4ma adj

Advertising
background image

7 Regular Maintenance Mode

GD-70D

- 58 -

↓ ↑

● ○ ○

SET 15

MNT OUT

MAINTENANCE

SET

2.5mA

MNT OUT

MAINTENANCE

↓↑

4.0mA

MNT OUT

MAINTENANCE

↓↑

HOLD

MNT OUT

MAINTENANCE

↓↑

4-20mA

MNT OUT

MAINTENANCE

SET 15. MNT OUT

Set the external output for the
maintenance mode.
Select either 2.5 mA/4.0
mA/HOLD (previous value)/4
- 20 mA (linked to display
value), and then press the
SET key to confirm the
selection.
(4 - 20 mA specification)

| ↑

| |

| |
| |

| |

| |
| |
| |

↓ |

● ○ ○

SET 16

MA 4-20

MAINTENANCE

SET

100%

4MA ADJ

MAINTENANCE

100%

20MA ADJ

MAINTENANCE

SET 16.MA 4-20

Adjust the external output (4 -
20 mA).
Adjust the output (%) by
pressing the ▲ or ▼ key, and
then press the SET key to
confirm the value.
(It must be adjusted to the
upper unit or the ammeter.)
After adjustment of 4 mA is
completed, perform the
adjustment of 20 mA.
(Press the MODE key to skip
this menu.)
(4 - 20 mA specification)

| ↑

| |

| |

| |

↓ |

● ○ ○

SET 17

BK LIGHT

MAINTENANCE

SET

ON

BK LIGHT

MAINTENANCE

↓↑

SAVE

BK LIGHT

MAINTENANCE

SET 17.BK LIGHT

Set the backlight.
Select either ON/SAVE, and
then press the SET key to
confirm the selection.
When ON is selected, the
backlight lights up
continuously.
When SAVE is selected, the
backlight is usually turned off,
but lights up only during an
operation or event.

| ↑

| |

↓ |

● ○ ○

SET 18

ETHERNET

MAINTENANCE

SET 18. ETHERNET

This is a setting screen of the
ETHERNET setting. This is
not used for the detector.

↓ ↑