1ppm, 0ppm, Zero – RKI Instruments GD-70D Series User Manual

Page 46: Zero set, Zero ok, Zero ng

Advertising
background image

6 User Mode

GD-70D

- 46 -

<Zero Adjustment "1-1">

This is used to perform the zero adjustment. Before starting the zero adjustment, let the detector suck the
zero adjustment gas and wait until the indicator is stabilized.
For oxygen deficiency alarm specification (OSU - 0 to 25 vol%), "1-1" is the span adjustment. In this case,
the AIR adjustment is performed, so that fresh air must be introduced to adjust it to 20.9 vol%. For
information on the span adjustment, see "7-2. How to perform a gas calibration".

Zero Adjustment

PW AL1

AL2

F

LCD

1-1.ZERO

Press the SET key.

● ○

1- 1

ZERO

MAINTENANCE

Current Concentration Value
Display

Press the SET key to perform the
zero adjustment.

● ○

1.1ppm

ZERO SET

MAINTENANCE

Zero Adjustment Completed

The menu returns to 1-1.ZERO
automatically.

● ○

0.0ppm

ZERO OK

MAINTENANCE

Return to 1-1.ZERO

* When the zero adjustment fails

● ○

0.0ppm

ZERO NG

MAINTENANCE

Return to 1-1.ZERO